Description
Detection Method and Diagram
- Collect the secretion samples from the vaginal wall (middle section) of a patient with a cotton swab (as many as possible). Take out a test tube and unscrew its cap.
- Dip the sampling cotton swab into the test tube solution, and stir and shake it properly.
- Put the test tube with the cotton swab in a thermostat at 37℃ for 10 min.
- Take out the chromogenic reagent bottle and unscrew its cap. Then, add one or two dropschromogenicreagent to the test tube containing a cotton swab, and shake it properly.
- Observe the color immediately within 3 min after the chromogenic reagent is added.
Result Interpretation (diagram):
a) Positive result: A blue or green color in the test tube or on the cotton swab indicates a positive result.
B) Negative results: A yellow color in the test tube indicates a negative result.
Storage and Stability
The BV RAPID BLUE™ detection kit shall be stored at 4℃-26℃ and used before the expiration date.
Result Reliability
Sensitivity: 92.7% Specificity: 96.9% Total Accuracy: 95.8%
| Step①
Collect vaginal secretions with a cotton swab |
Step②
Insert the cotton swab into a test tube |
Step③
Put the test tube in a thermostat at 37℃ for 10min |
Step④
Add two drops of chromogenic agent |
Step⑤ | |
| Negative | Positive | ||||

Reviews
There are no reviews yet.